Center for Devices and Radiological Health
Final
_____________________________________________________________
DESIGN CONTROL REPORT
AND GUIDANCE
______________________________________________________________
Revised May 1998
All Previous Editions Are Obsolete
Effective starting on June 1, 1998
Instructions:
Notes:
2. A negative response to a DC question is not necessarily a citable deficiency. For example: Not all of the potential inputs listed in question 2 of "820.30(c) Design input" are necessary for all design projects.
82830D = Domestic Design Control Inspections
83830D = Domestic Pre-Approval Design Control Inspections
82R915 = Foreign Design Control Inspections
83R915 = Foreign Pre-Approval Design Control Inspections
Design Control Report
Name of Establishment:
CFN or FEI Number:
Date(s) of Inspection:
Type of Inspection: PMA [ ] Q.S. Reg. [ ] 510(k) [ ] Other [ ]
820.30(a) General
Regulatory Requirements
|
Section |
Device |
| 868.6810
878.4460 880.6760 892.5650 892.5740 |
Catheter, Tracheobronchial, Suction
Glove, Surgeon’s Restraint, Protective System, Applicator, Radionuclide, Manual Source, Radionuclide, Teletherapy |
Instructions
1. Select a device that was subject to design controls. Describe the device. What is it used for? Who will use it (e.g., doctors, nurses, patients, etc.)? Indicate whether this was a design project for an original design or a modification of an existing design with a Y for Yes or an N for No.
|
Description of Device |
|
|
What is the device used for? |
|
|
Who will use the device? |
|
|
Y or N |
|
| Original design | |
| Modification of existing design |
2. For the device selected, identify at what stage in the design and development effort design controls were applied. If the design and development effort has not been completed, identify the current status of the design and development effort. If the design and development effort was initiated prior to June 1, 1998, at what stage was the design process on June 1, 1998?
|
Activity or Date |
Stage of Design & Development |
| When design controls were applied | |
| Currently | |
| As of June 1, 1998 |
820.30(b) Design and development planning
Regulatory Requirements
Each manufacturer shall establish and maintain plans that describe or reference the design and development activities and define responsibility for implementation. The plans shall identify and describe interfaces with different groups or activities that provide, or result in, input to the design and development process. The plans shall be reviewed, updated, and approved as design and development evolves.
Instructions
|
Responsibility Assigned? Y or N |
Design Activity |
Alternative Terminology |
| Risk analysis | ||
| Design input | ||
| Design output | ||
| Design review | ||
| Design verification | ||
| Design validation | ||
| Design transfer | ||
| Design changes | ||
| Interfaces |
820.30(c) Design input
Regulatory Requirements
Each manufacturer shall establish and maintain procedures to ensure that the design requirements relating to a device are appropriate and address the intended use of the device, including the needs of the user and patient. The procedures shall include a mechanism for addressing incomplete, ambiguous, or conflicting requirements. The design input requirements shall be documented and shall be reviewed and approved by a designated individual(s). The approval, including the date and signature of the individual(s) approving the requirements, shall be documented.
Instructions
|
Y, N or NA |
Potential Design Input |
| Intended use | |
| User/patient/clinical | |
| Performance characteristics | |
| Safety | |
| Limits and tolerances | |
| Risk analysis | |
| Toxicity and biocompatibility | |
| Electromagnetic compatibility (EMC) | |
| Compatibility with accessories/auxiliary devices | |
| Compatibility with the environment of intended use | |
| Human factors | |
| Physical/chemical characteristics | |
| Labeling/packaging | |
| Reliability | |
| Statutory and regulatory requirements | |
| Voluntary standards | |
| Manufacturing processes | |
| Sterility | |
| MDRs/complaints/failures and other historical data | |
| Design history files (DHFs) | |
| Other (describe): |
3. Review the process for resolving incomplete, ambiguous, or conflicting requirements. For the design reviewed, determine whether any incomplete, ambiguous, or conflicting requirements were not resolved per the manufacturer’s procedures.
4. Review how the design input addresses the user interface: the hardware (and software, if applicable) and features that define the interactions between users and the device. Examples are exploratory studies (e.g., interviews), usability studies (e.g., user evaluation, task analysis, risk analysis, or workload analysis), or any combination thereof.
5. Confirm that design input is reviewed, approved and documented.
820.30(d) Design output
Regulatory Requirements
Each manufacturer shall establish and maintain procedures for defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements. Design output procedures shall contain or make reference to acceptance criteria and shall ensure that those design outputs that are essential for the proper functioning of the device are identified. Design output shall be documented, reviewed, and approved before release. The approval, including the date and signature of the individual(s) approving the output, shall be documented.
Instructions
4. Confirm that design output is reviewed, approved and documented.
820.30(e) Design review
Regulatory Requirements
Each manufacturer shall establish and maintain procedures to ensure that formal documented reviews of the design results are planned and conducted at appropriate stages of the device’s design development. The procedures shall ensure that participants at each design review include representatives of all functions concerned with the design stage being reviewed and an individual(s) who does not have direct responsibility for the design stage being reviewed, as well as any specialists needed. The results of a design review, including identification of the design, the date, and the individual(s) performing the review, shall be documented in the design history file (the DHF).
Instructions
820.30(f) Design verification
Regulatory Requirements
Each manufacturer shall establish and maintain procedures for verifying the device design. Design verification shall confirm that the design output meets the design input requirements. The results of the design verification, including identification of the design, method(s), the date, and the individual(s) performing the verification, shall be documented in the DHF.
Instructions
820.30(g) Design validation
Regulatory Requirements
Each manufacturer shall establish and maintain procedures for validating the device design. Design validation shall be performed under defined operating conditions on initial production units, lots, or batches, or their equivalents. Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation, shall be documented in the DHF.
Instructions
|
|
| Y or N |
Design Validation Activity |
| Clinical studies approved via Institutional Review Boards (IRBs) and Investigational Device Exemptions (IDEs), or IRB approvals alone for non-significant (risk) devices. | |
| Clinical evaluations in clinical or non-clinical settings | |
| Simulated use (conditions) | |
| 510(k) historical database search | |
| Literature search | |
| Review of labels and labeling, packaging, and other historical product information | |
| Other (describe): |
4. If the chosen device contains software, review the software validation.
|
Y or N |
Risk Analysis Tool/Technique |
Standard used (if applicable) |
| Failure Mode Effects Analysis (FMEA) | ||
| Failure Mode Effects and Criticality Analysis (FMECA) | ||
| Fault Tree Analysis (FTA) | ||
| In-House | ||
| Other (describe): |
6. Confirm that design validation data show the approved design meets the predetermined user needs and intended use(s) and that the completed design validation did not leave any unresolved discrepancies.
820.30(h) Design transfer
Regulatory Requirements
Each manufacturer shall establish and maintain procedures to ensure that the device design is correctly translated into production specifications.
Instructions
820.30(i) Design changes
Regulatory Requirements
Each manufacturer shall establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation.
Instructions
Note: 820.30(i) relates to Product Design Changes. Changes to existing Production Processes should also be evaluated under 820.70(b).
5. Confirm design changes are reviewed, approved and documented.
820.30(j) Design history file
Regulatory Requirements
Each manufacturer shall establish and maintain a DHF for each type of device. The DHF shall contain or reference the records necessary to demonstrate that the design was developed in accordance with the approved design plan and the requirements of this part.
Instructions
Design Control Objectionable Conditions and Discussions with Management:
|
|
|
Name(s) |
|
|
Title(s) |
|
| Signature(s) | |
| Date |