This document is intended to provide guidance in the preparation of a
regulatory submission.
It does not bind the FDA or the regulated industry in any manner.
Office of Device Evaluation
Division of Cardiovascular, Respiratory, and Neurological Devices
Circulatory Support and Prosthetic Devices Group
Document issued on: March 10, 1997
While this guidance document represents a final document, comments and suggestions may be submitted at any time for Agency consideration by writing to:
Steven Allis (HFZ-450)
Food and Drug Administration
9200 Corporate Boulevard
Rockville, Maryland 20850.
For questions regarding the use or interpretation of this guidance, contact Steven Allis
by phone at 301-443-8262
U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES
Public Health Service
Food and Drug Administration
Center for Devices and Radiological Health
Scope
This guidance is intended to aid in the preparation or review of premarket notification
(510(k)) applications for some of the devices regulated under:
CFR Section: 21 CFR §870.1130, Noninvasive Blood Pressure Measurement System
Class: II
Panel: Circulatory System Devices Panel (74)
Product Code: DXN
This guidance applies to non-invasive blood pressure (NIBP) monitors covered by the
ANSI/AAMI SP10-1992 standard for electronic or automated sphygmomanometers (SP10
standard). Included in the SP10 standard are automated NIBP monitors which measure blood
pressure at the arm, wrist, or finger of the patient using a standard oscillometric
measurement method.
This guidance does not apply to NIBP monitors excluded by the SP10 standard, and those
which use a non-oscillometric (or non-standard oscillometric) measurement method. Despite
this limitation, the information contained in this guidance may be helpful to any NIBP
monitor application.
This guidance is complementary to the requirements of 21 CFR § 807.87. Other
information not identified in this guidance may be required in a 510(k) application. This
guidance is subordinate to all other applicable statutes, regulations, and policies.
Recommended Information and Testing
1. Device Description
The description should include sufficient information to define the design,
capabilities, and function of the device, and the scope of the 510(k) submission. Minimal
information includes:
If the 510(k) application is for a modification to an existing device, the manufacturer
should provide the specifications for the original device and a detailed and complete
description of the similarities and differences between the two versions of the device.
Lastly, the manufacturer should provide a table comparing the predicate device to the new
device for the items listed above.
2. In-Vitro and Clinical Performance Testing
Substantial equivalence can be demonstrated by showing either 1) sufficient comparison
testing with a legally-marketed predicate device, 2) conformance to the SP10 standard, or
3) conformance to any foreign or domestic standard which meets or exceeds the requirements
of the SP10 standard.
Comparison Testing
It is strongly recommended that substantial equivalence be demonstrated by showing
conformance to the SP10 standard. However, if the manufacturer chooses to provide
comparative testing, the provided data should meet the General Requirements (listed
below) and account for the following:
SP10 Standard Testing
To show conformance to the SP10 standard, the manufacturer should list each of the
requirements of the standard and describe how the device conforms to each requirement. For
every requirement which necessitates clinical or in-vitro testing, the test
protocol, test data and results, and analysis should be provided and clearly identified.
The necessary detail for each element of a test report is described below in General
Requirements. For devices with unique features or intended uses, additional testing
beyond the SP10 standard may be necessary.
If the SP10 standard is chosen by the manufacturer, conformance to the entire standard
is necessary. Conformance to portions of the standard is insufficient to permit the
standard's use or to allow a labeling claim to that effect. Therefore, if only part of the
standard is met, the manufacturer should refer to the Comparison Testing section of
this guidance.
Foreign Standards
If the manufacturer chooses to conform to a standard other than the SP10, it is
recommended that they list each requirement of the SP10 standard, compare the foreign
standard to the SP10 requirements, and clearly identify where the foreign standard does
not meet the requirements of the SP10 standard (if at all). Justification for any
differences should be based on valid scientific or statistical analyses and supported by
testing if necessary.
Additional Testing
In addition to the testing described above, it is necessary that the manufacturer
evaluated the following:
General Requirements
In general, any in-vitro test report should include the following:
In general, any clinical data necessitates consideration or inclusion of the following:
3. In-Vitro Safety Testing
Environmental Testing
The manufacturer should evaluate the ability of the device to function after exposure
to the environmental hazards expected when used by an abusive user. Tests for some of
these hazards may be found in SP10, IEC 601-1, and IEC 529. These hazards minimally
include:
Software
To demonstrate the quality of the software used in or with the device, the following is
necessary:
Electrical Safety
Any appropriate standard for electrical safety may be used. If the SP10 standard is
used, the manufacturer should conform to the standard's requirements or justify a
modification to the standard.
Electromagnetic Compatibility
Electromagnetic compatibility (EMC) testing is necessary to demonstrate that the device
(1) will not adversely interfere with the performance of other electronic devices
(emissions), and (2) will perform as expected in the presence of other electronic devices
or other sources of electromagnetic interference (EMI) in the intended environment of use
(immunity). To demonstrate EMC for the device, the following information is necessary:
Any omitted tests or deviations from the requirements of the chosen standard require
justification. In addition, the manufacturer should provide a list of all known or
suspected EMI incidents associated with the device, the results of any related
investigations, a description of any corrective action taken, and any device labeling that
references EMC or EMI.
Biocompatibility
Identification of all patient and operator contacting materials and conformance to the
International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1:
Evaluation and Testing," is necessary. For materials that are widely used in the same
or similar applications, supportive information demonstrating the material's use in other
medical devices or products may be acceptable in establishing biocompatibility. However,
any references should be to the same vendor and material, and account for any changes to
the material due to subsequent processes or manufacturing (e.g., sterilization, forming,
melting).
Sterilization
Usually, there are no sterile components to NIBP monitors. If sterile components are
identified, refer to the 510(k) Sterility Review Guidance #K90-1 (February 1990) or
the most recent sterilization policy or guidance.
Packaging
The manufacturer should describe all packing for the device. This description should
include a description of the design, materials, and the sealing method.
For package integrity, any appropriate standard may be used. If the SP10 standard is
used, the manufacturer should conform to the standard's requirements or justify any
modification.
Shelf Life
Usually a shelf life is not necessary for NIBP monitors. However, if the device
contains any sterile or degradable components, shelf life data may be necessary.
4. Labeling
Conformance to the labeling regulations and policies is necessary. Appropriate labeling guidances are available through the Division of Small Manufacturers Assistance (DSMA) at its toll-free number (800) 6382041 .
If the SP10 standard is used, the labeling requirements of the standard should be
included or justification provided for any modifications.
5. Regulatory Requirements
Either a Summary of Safety and Effectiveness or a 510(k) Statement is necessary as
described in 21 CFR § 807.92 and 21 CFR § 807.93, respectively.
A "Truthful and Accuracy Statement" is necessary according to 21 CFR §
807.87 (j).
An "Indications for Use Statement" is necessary according to Office of Device
Evaluation policy. A format for this statement can be provided to the manufacturer by
DSMA.
(Updated April 15, 1997)